Продукция
Портативный фетальный монитор YSFM100 для матерей 10,4 дюйма
Предмет №: YSFM100
10,4-дюймовые Материнские портативные Фетальные Мониторы YSFM100, лидер продаж
ОПИСАНИЕ
High Quality 10.4 Inches Maternal Portable Fetal Monitor YSFM100 Hot Sale
YSFM100 ProductsYSFM100 Specification
Physical Specifications | ||||
Dimension and Weight | Dimension | 32cm (W)*26cm (D)*8cm (H) | ||
Weight | standard configuration: 2.8 kg | |||
Operation Environment | Power requirements | AC100-240V , 50/60Hz | ||
Power | 50VA | |||
Temperature | 0-40℃ | |||
Humidity | 0-93% non-condensing | |||
Fetal/Maternal Performance Specifications | ||||
Performance Specifications | Display | 10.4 inches (1024×768 pixel) color TFT –LCD | ||
Indicator | Visual on-display and integrated, alarm light, audible alarm | |||
Battery | Internal rechargeable Li-ion battery power | |||
Alarm | Visual on-display and integrated, alarm light, audible alarm | |||
Memory | 24-hours memory for fetal traces 150 groups patient data |
|||
Printer | America standard/international standard Built-in thermal printer (paper width:110mm) 1/2/3cm/min real-time printing speed, fast print speed (stored traces) |
|||
Paper | Z-fold, thermosensitive | |||
Record Message | ID, name, age, date, time, print speed etc | |||
CMS | Support connecting to central monitoring station through LAN (TCP/IP) or wireless connection | |||
Fetal/Maternal Technical Specifications | ||||
Ultrasound (FHR) | Measurement range | US | 30 to 240 bpm | |
Resolution | Display | 1bpm | ||
Printer | 1bpm | |||
jitter@200 bpm | ≤3bpm | |||
Display update rate | 0.25second | |||
Beat to bea change(max.) for ultrasound | 28bpm | |||
US frequency | 1MHz±5%Hz | |||
TOCO | Measurement method | Strain gauge sensor element | ||
Sensitivity | 1 unit =2.5g | |||
Resolution | 1 unit | |||
Measurement range | 500 units | |||
Baseline setting | 0, 5,10, 15, 20 can be selective | |||
Update rate | 0.25 second | |||
Auto offset correction | 3 second after connecting the transducer | |||
Auto zero adjust | TOCO value is set to zero negative following a negative Measure value for 5 seconds |
|||
Auto Fetal Movement (AFM) | Technique | Pulsed doppler ultrasound | ||
Range | 0-100% | |||
Resolution | 1% | |||
Automatic fetal movement actograph | ||||
Manual fetal movement marker | ||||
Maternal NIBP(MNIBP) | Operation mode | Manual/automatic/ STAT | ||
Measurement unit | mmHg / kPa selectable | |||
Test interval | 0-480 minutes | |||
Accuracy | Max.Std.deviation:8 mmHg(1.1kPa) | |||
Max. mean Error: ±5mmHg(±0.7kPa) | ||||
Measurement range | Systolic | 30 to 270 mmHg(4 to 36 kPa) | ||
Diastolic | 10 to 245 mmHg(1.5 to 32kPa) | |||
Mean | 20 to 255 mmHg(2.5 to 34kPa) | |||
Pulse rate range | 20-250bpm | |||
Over-pressure protection | Yes | |||
Maternal SpO2 (MSpO2) | Measurement range | 0%-100% | ||
Resolution | 1% | |||
Accuracy | ±2% | |||
Pulse rate range accuracy | ±3bpm |
High-tech fetal& maternal monitor,it offers a
reliable, high-performance solution for fetal monitoring needs in the clinic
and hospital, from pre-natal check-ups to antepartum monitoring of high-risk
pregnancies.
YSFM100 K Features:
♦ Folded 10.4 inches TFT LCD high-brightness screen
♦ 12-crystal waterproof pulse Doppler transducer (optional), providing wider beam area and more homogeneous signal, thus realizing better performance for beside monitors than conventional transducers.
♦ Event Marker makes it easy to annotate perceived fetal movements (FM),
or other events on the strip chart.
♦ Fetal Movement Profile(FMP) detects and automatically records gross fetal body movements.
♦ Heart rate offset mode makes it easy to visually distinguish between the heart rates of twins by allowing you to offset the secondary FHR + 30BPM.
♦ Heartbeat Coincidence recognition provides visual and audible indication when it detects synchronous fetal or maternal heart-rate signals, indicating that you may be monitoring a duplicate signal.
♦ Fetal heart rate alarm provides both audible and visual alerts if the heart rate is outside of the user-defined high/low range.
♦ Easy-to-load Z-fold paper stacks itself neatly upon exiting the recorder, facilitating easy storage.
♦ Internal Li-on battery power or AC power from external power supply.
♦ Supporting TCP/IP or wireless connection to the central monitoring station
♦ NST Report (Fetus health report)
Fetal Acoustic Stimulator(FASt)
The optional Fetal Acoustic Stimulator uses low frequency vibro-acoustic stimuli(VAS) to “wake” the fetus, increasing the efficiency of antepartum testing and improving retest consistency. VAS indicated with a musical note on the strip
YOUR MAY ALSO LIKE ...