We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.

Professional Medical Equipment Supplier - Yuesen Med Professional Medical Equipment Supplier - Yuesen Med

Hot Products

X Ray Machine  To  Zimbabwe

x ray machine, 50KW x ray machine

Microscope To Malawi

Achromatic objectives: 4X、10X、40X(S), 100X(S、Oil) Wide field eyepiece: WF10X(WF16X for option) Eyepiece head: Sliding binocular head inclined at 45° Stage: Double layer mechanical stage size 140X140mm, moving range 75X45mm Focusing: Coaxial coarse and

Color Doppler Trolley System To Somalia

Color Doppler Trolley System To Somalia

Automated Hematology Analyzer To Peru

High quality Automated Hematology Analyzer (3-part differential) with cheap price

The several faults of Ultrasound probe

Views : 967
Author : Monica Liang
Update time : 2015-12-21 08:00:00

Ultrasound diagnosis equipment uses ultrasonic to extract all kinds of information of human tissue, such as organization interface information, blood flow information, so ultrasound probe, which can transfer information. Ultrasound probe is the key part of ultrasound diagnosis and its value has accounted for a quarter to a third of one ultrasound instrument total value, like B ultrasound probe is the same. In addition, doctors use it to check patients and it is easy to have faults. Therefore, it is necessary for reasonable using and maintenance.

In general, there are several faults of ultrasound probe as follows: Appearance cracking, internal sparkling, sensitivity decreasing, cable cracking and electrode lead break.

 First, Appearance cracking, internal sparkling, reduced sensitivity.

(1)Although ultrasound probe manufacturers are different, the basic structure is the same in general. Put the transducer in probe cavity of plastic wrap.

 As we know, when we examinate, checking parts have to been embrocated enough lotus root mixture, but this lotus root mixture is the biggest threat of plastic wrap. In most conditions, probe surface lotus root mixture is not clear and cause appearance cracking after finishing operation. After a long time, lotus root mixture has corrosion for the plastic and rubber band and damage probe. Therefore we must clean the surface lotus root mixture after finishing work and you had better use clean and wet soft cloth to wipe it; The other reason of appearance cracking is that probe is often influenced by the sun and large temperature difference. Because the sun will make plastic materials early ageing and deformation. So, ultrasound equipment should be placed in constant greenhouse and avoid light ray. For example probe frequency instability, you can consider probe moisture, then there is no effect after using no water copper sulfate moisture absorption; The following you use special tool to open probe shell and take down the circuit board and find that there is more circuit board oil and use 99.5% pure alcohol to soak ten minutes; Then use fuzz wipe oil and dirty; The next step, you can use fan to blow dry and put it back and open the machine, probe fault is removed.

(2) Internal sparkling. In general, probe production process defect causes sensitivity decrasing. It will happened after a period of time, especially bad penetration ability after adjusting host parameters.

Second, Circle image causes bad local

In general, chip problem cause this fault. For example, there is a white bright field from up to down and there is no imge information in this field. After checking, the probe internal chip damage causes this.

Third, Cable craze or electrode lead break

Due to probe move back and forth in use process, it will pull cable automatically, especially in turning probe process, it will make probe and the cable wind. Therefore, bad operation habit or cable placed bad cause this fault. For this fault, you can open probe shell and rearrange all lines and weld well, then use super glue sticky shell well.

 
Related News
Read More >>
Operation video of YSENMED YSDXA-2 Dual-Energy X-ray Absorptiometry Bone Densitometry. Operation video of YSENMED YSDXA-2 Dual-Energy X-ray Absorptiometry Bone Densitometry.
Nov .22.2024
Here we share the operation video of YSENMED YSDXA-2 Dual-Energy X-ray Absorptiometry Bone Densitometry.
Australian vet clinic digitizing its x-ray with YSFPD-M1717V Vet DR panel Australian vet clinic digitizing its x-ray with YSFPD-M1717V Vet DR panel
Nov .21.2024
Australian vet clinic upgrades its analog x-ray system by bringing in YSENMED YSFPD-M1717V Vet wireless flat panel detector.
Bolivian doctor satisfied with YSB-C10RL USB & WIFI color ultrasound probe and considers re-purchasing Bolivian doctor satisfied with YSB-C10RL USB & WIFI color ultrasound probe and considers re-purchasing
Nov .20.2024
Bolivian doctor is satisfied with our USB & WIFI color doppler ultrasound probe YSB-C10RL, and considers re-purchasing another two pieces.
Two veterinary hematology analyzers YSTE320V starting to serve in a vet clinic in Honduras. Two veterinary hematology analyzers YSTE320V starting to serve in a vet clinic in Honduras.
Nov .19.2024
Two veterinary hematology analyzers have arriaved at a vet clinic in Honduras and the vet is happy with them.